Measurement and Instrumentation Kelvin’s double bridge is used to measure low resistance because effect of contact and lead resistance is eliminated. there is no thermo electric emf. it has high sensitivity. resistance variation due to temperature. effect of contact and lead resistance is eliminated. there is no thermo electric emf. it has high sensitivity. resistance variation due to temperature. ANSWER DOWNLOAD EXAMIANS APP
Measurement and Instrumentation A permanent magnet moving coil ammeter is connected in 50 Hz a. c. Circuit in which 5A current is flowing. The meter will read 0 A None of the listed here 2.5 A 5 A 0 A None of the listed here 2.5 A 5 A ANSWER DOWNLOAD EXAMIANS APP
Measurement and Instrumentation The total current I = I1 + I2 in a circuit is measured as I1 = 150 ± 1A , I2 = 250 ± 2A, where the limits of error are as standard deviations. I is measured as (400 ± 1) A (400 ± 1.5) A (400 ± 2.24) A (400 ± 3) A (400 ± 1) A (400 ± 1.5) A (400 ± 2.24) A (400 ± 3) A ANSWER DOWNLOAD EXAMIANS APP
Measurement and Instrumentation A 50 Hz bar primary CT has a secondary with 500 turns. The secondary supplies 5 A current into a purely resistive burden of 1Ω. The magnetizing ampere-turn is 200. The phase angle between the primary and secondary current is 175.4o. 4.6o. 85.4o. 94.6o. 175.4o. 4.6o. 85.4o. 94.6o. ANSWER DOWNLOAD EXAMIANS APP
Measurement and Instrumentation Most sensitive galvanometer is Elastic galvanometer Vibration galvanometer Spot ballastic galvanometer Duddle galvanometer Elastic galvanometer Vibration galvanometer Spot ballastic galvanometer Duddle galvanometer ANSWER DOWNLOAD EXAMIANS APP
Measurement and Instrumentation It can be stated that CT operates with considerably lower flux density than a PT CT operates at almost the same flux density as a PT CT operates with a higher flux density than PT No generalisation can be made with regard to the flux densities in CT and PT CT operates with considerably lower flux density than a PT CT operates at almost the same flux density as a PT CT operates with a higher flux density than PT No generalisation can be made with regard to the flux densities in CT and PT ANSWER DOWNLOAD EXAMIANS APP